Voltage DC |
110.000 mV |
0.025%+0.015% * |
1.10000 V |
0.025%+0.005% * |
11.0000 V |
0.025%+0.005% * |
110.000 V |
0.05%+0.005% * |
300.00 V |
0.05%+0.005% * |
Voltage AC |
20 to 40 Hz |
2% + 10 |
40 to 500 Hz |
0.5% + 5 |
500 to 1 kHz |
2% + 10 |
1 kHz to 5 kHz |
10% + 20 |
Ranges: |
1.1000, 11.000, 110.00, 300V |
Current DC |
30.000 mA |
0.01% + 0.015% * |
110.00 mA |
0.01% + 0.015% * |
Resistance |
11.000 Ω |
0.05% + 50 mΩ |
110.00 Ω |
0.05% + 50 mΩ |
1.1000 kΩ |
0.05% + 500 mΩ |
11.000 kΩ |
0.1% + 10 Ω |
Frequency |
1.00 to 109.99 Hz |
0.05 Hz |
110.0 to 1099.9 Hz |
0.5 Hz |
1.100 to 10.999 kHz |
5 Hz |
11.00 to 50.00 kHz |
50 Hz |
Pressure |
Accuracy |
from 0.025% of range using any of 29 pressure modules. |
|
(for detailed specifications refer to pressure modules in options and accessories) |
|
Modules available for differential, gauge, vacuum, absolute, dual and high pressure. |
Note |
* |
(% of reading + % of full scale) |
Source Accuracy |
Voltage DC |
110.000 mV |
0.01%+0.005% * |
1.10000 V |
0.01%+0.005% * |
15.0000 V |
0.01%+0.005% * |
Current DC |
22.000 mA (Source) |
0.01%+ 0.015% * |
22.000 mA (Simulate) |
0.02% + 0.03% * |
Resistance |
11.000 Ω |
0.01% + 20 mΩ |
110.00 Ω |
0.01% + 40 mΩ |
1.1000 kΩ |
0.02% + 500 mΩ |
11.000 kΩ |
0.03% + 5Ω |
Frequency |
0.00 to 10.99 Hz |
0.01 Hz |
11.00 to 109.99 Hz |
0.1 Hz |
110.0 to 1099.9 Hz |
0.1 Hz |
1100 to 21999 Hz |
2 Hz |
22.000 to 50.000 kHz |
5 Hz |
Note |
* |
(% of reading + % of full scale) |
RTDs and Thermocouples |
Measure Accuracy: |
10 Ω Cu (427): |
2 °C |
100 Ω Pt (3916): |
0.3 °C |
100 Ω Pt (3926): |
0.3 °C |
100 Ω Pt (385): |
0.3 °C |
200 Ω Pt (385): |
0.3 °C |
500 Ω Pt (385): |
0.3 °C |
1000 Ω Pt (385): |
0.3 °C |
120 Ω Ni (672): |
0.3 °C |
note: |
For 2-and 3-wire measurement, add 0.4 °C |
E: |
0.3 °C |
N: |
0.5 °C |
J: |
0.3 °C |
L: |
0.3 °C |
K: |
0.3 °C |
T: |
0.3 °C |
U: |
0.3 °C |
B: |
0.9 °C |
R: |
1.0 °C |
S: |
0.9 °C |
C: |
0.6 °C |
note: |
Accuracy with external cold junction, for internal junction add 0.2°C |
Source Accuracy: |
10 Ω Cu (427): |
1 °C |
100 Ω Pt (3916): |
0.1 °C |
100 Ω Pt (3926): |
0.1 °C |
100 Ω Pt (385): |
0.1 °C |
200 Ω Pt (385): |
0.1 °C |
500 Ω Pt (385): |
0.1 °C |
1000 Ω Pt (385): |
0.1 °C |
120 Ω Ni (672): |
0.1 °C |
note: |
For 2-and 3-wire simulation, add 0.4 °C |
E: |
0.2 °C |
N: |
0.3 °C |
J: |
0.2 °C |
L: |
0.2 °C |
K: |
0.3 °C |
T: |
0.3 °C |
U: |
0.3 °C |
B: |
0.8 °C |
R: |
0.9 °C |
S: |
0.9 °C |
C: |
0.6 °C |
note: |
Accuracy with external cold junction, for internal junction add 0.2 °C |
Specifications |
Ramp Functions |
Source functions: |
Voltage, current, resistance, frequency, temperature |
Rate: |
4 steps/second |
Trip detect: |
Continuity or voltage (continuity detection not available when sourcing current) |
Loop Power Function |
Voltage: |
Selectable, 24 V or 28 V |
Accuracy: |
5% |
Maximum current: |
22 mA, short circuit protected |
Maximum input voltage: |
30 V DC |
Step Functions |
Source Functions |
Voltage, current, resistance, frequency, temperature |
Manual Step |
Selectable step, change with arrow buttons |
Autostep |
Fully programmable for function, start delay, stepvalue, time per step, repeat |
Environmental Specifications |
Operating Temperature |
|
-10 °C to +50 °C |
|
-20 °C to +50 °C * |
Storage Temperature |
|
-20 °C to +60 °C |
Dust/water resistance |
|
Meets IP52, IEC 529 |
Operating Altitude |
|
2800 m |
Note |
* |
Except frequency and AC |
Safety Specifications |
Agency Approvals |
|
CAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993 |
Mechanical & General Specifications |
Size |
|
130 x 236 x 61 mm |
Weight |
|
1.4 kg |
Batteries |
NiCd: |
7.2V, 1.7 Ah |
Battery Life |
|
~8 hours typical |
Battery Replacement |
|
Via snap-shut door without opening calibrator; no tools required |
Side port connections |
|
Pressure module connector |
|
Connection for optional battery eliminator |
Data storage capacity |
|
1 day of calibration results |
90 day specifications |
|
The standard specification intervals for the 741B is 1 and 2 years. |
|
Typical 90 day measurement and source accuracy can be estimated by |
|
dividing the one year "% of reading" or "%of output" specifications by 2. |
|
Floor specifications, expressed as "% of full scale" or |
|
"counts" or "ohms" remain constant. |